Fluke 101 Basic Digital Multimeter
|
Electrical Specifications |
|
V AC (40-500Hz) |
Range |
600.0 mV
6.000 V
60.00 V
600.0 V |
|
Resolution |
0.1 mV
0.001 V
0.01 V
0.1 V |
|
Accuracy |
3.0%±3 digit
1.0%±3 digit |
|
V DC |
Range |
6.000 V
60.00 V
600.0 V |
|
Resolution |
0.001 V
0.01 V
0.1 V |
|
Accuracy |
0.5%±3 digit |
|
Resistance Ohms |
Range |
400.0 ?
4.000 k?
40.00 k?
400.0 k?
4.000 M?
40.00 M? |
|
Resolution |
0.1 ?
0.001 k?
0.01 k?
0.1 k?
0.001 M?
0.01 M? |
|
Accuracy |
0.5 %±3 digit
0.5 %±2 digit
0.5 %±2 digit
0.5 %±2 digit
0.5 %±2 digit
1.5%±3 digit |
|
Capacitance |
Range |
50.00 nF
500.0 nF
5.000 µF
50.00 µF
100.0 µF |
|
Resolution |
0.01 nF
0.1 nF
0.001 µF
0.01 µF
0.1 µF |
Frequency
(10 Hz – 100 kHz) |
Range |
50.00 Hz
500.0 Hz
5.000 kHz
50.00 kHz
100.0 kHz |
|
Resolution |
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz |
|
Duty Cycle |
Range |
0.1 % to 99.9 % |
|
Resolution |
0.1 % |
|
General Specifications |
|
CAT Rating |
CAT III 600 V |
|
V AC Range |
600.0 V |
|
V DC Range |
600.0 V |
|
Ohms Range |
40.00 M? |
|
Auto Shutoff |
Y |
|
Continuity |
Y |
|
Capacitor |
100.0 µF |
|
Frequency |
100.0 kHz |
|
Hold |
No |
|
Duty Cycle |
Yes |
|
Size |
130 mm x 65 mm x 27mm |
|
Weight |
160 g |
|
Battery |
Two AAA |
|
Warranty |
1 year |
|
Environmental Specifications |
|
Operating Temperature |
0 °C to +40 °C |
|
Storage Temperature |
-30 °C to +60 °C |
|
Operating Humidity |
Non condensing (<10 °C)
<=90% RH (at 10 °C to 30 °C)
<=75% RH (at 30 °C to 40 °C)
(Without Condensation) |
|
Operating Humidity, 40 M? |
80% RH 10 °C to 30 °C, 70% RH 30 °C to 40 °C |
|
Operating Altitude |
2,000 meters |
|
Storage Altitude |
12,000 meters |
|
IP Rating |
IP 40 per IEC 60529 |
|
Vibration Requirements |
MIL-PRF-28800F, Class 2 |
|
Drop Test Requirements |
-10°C and 1 meter to surface, per IEC-61010-1, and Fluke SOP 39.1 for portable Hand-held equipment
Shipping container drop per Fluke SOP 39. |
|
EMI, RFI, EMC |
Must meet all applicable requirements in
IEC/EN 61326-1 |
|
Absorption/Corrosion |
Per Fluke SOP 39.1 |
|
Temperature Coefficients |
Add 0.1 x specified accuracy for each degree C above 28 °C or below 18 ºC |
Fluke i400E AC Current Clamp
|
Electrical specifications |
|
AC current range |
1 A to 400 A
|
|
Output signal 1 |
mV/A ac |
|
Accuracy |
2.0 % ± 5 digits (45 Hz to 65 Hz)
2.5 % ± 5 digits (65 Hz to 400 Hz) |
|
Typical bandwith |
5 Hz to 20 kHz |
|
Working voltage |
600 V ac rms, in compliance with IEC 61010-1 and IEC 61010-2-030 |
|
Load impedance |
1 A to 400 A |
|
Maximum non-destructive current |
600 A |
|
Jaw inner diameter |
30 mm |
|
Maximum wire size |
600 MCM |
|
Standard plugs |
19.04 mm dual banana plugs |
|
Safety specifications |
|
Category rating |
IEC 61010-1 and IEC 61010-2-030/IEC 61010-2-031 and
IEC 61010-2-032: 600 V CAT III
Pollution Degree 2 |
|
Electromagnetic environment |
IEC 61326-1: Portable |
|
General Specifications |
|
Output cable length |
1 m |
|
Operating temperature |
-10 °C to +50 °C |
|
Storage temperature |
- 30 ºC to +60 ºC |
|
Relative humidity |
=90 % at 10 °C to 530 °C;
=75 % at 30 °C to 540 °C;
=45 % at 40 °C to 50 °C |
|
IP rating |
IEC 60529: IP30 when jaw closed |
|
Temperature coefficients |
Add 0.1 x specified accuracy for each °C above 28°C or below 18 °C. |
|
Operating altitude |
2,000 meters |
|
Storage altitude |
12,000 meters |
|
Vibration requirements |
MIL-PRF-28800F, Class 2 |
|
Drop test requirements |
-10 °C, 1m, 6 sides, oak wood |
|
Jaw impact test requirements |
Follow EN/IEC 61010-2-032:2012 |
|
EMI, RFI, EMC |
IEC/EN 61326-1 |
|
Dimensions (hxwxd) |
146 mm x 62 mm x 34 mm |
|
Weight |
150 g |
|
|
|
|